Telah dilakukan kajian penentuan logam berat dengan atomic absorption spectrophotometer (AAS) menggunakan metode solid phase extraction (SPE). Hasil penelitian menunjukkan bahwa penentuan logam berat tembaga (Cu) kadmium (Cd), dan timbal (Pb) dengan AAS menggunakan metode solid phase extraction (SPE) dapat dilakukan dengan adsorben kitin terimobilisasi ditizon. Adsorbsi logam tembaga (Cu) dan cadmium (Cd) optimum pada pH maksimal 6 sedangkan adsorbsi logam timbal (Pb) optimum pada pH maksimal 5. Penentuan logam tembaga (Cu) menggunakan metode solid phase extraction (SPE) dapat optimum dengan faktor pemekatan 4 kali dengan recovery 100 %, sedangkan logam kadmium (Cd) dapat optimum dengan faktor pemekatan 1,37 kali dan recovery 34 % dan timbal (Pb) optimum dengan faktor pemekatan 2,90 kali dan 72 %.
Kata kunci : logam berat, solid phase extraction (SPE), kitin terimobilisasi ditizon, atomic absorbtion spectrophotometer (AAS)
Kata kunci : logam berat, solid phase extraction (SPE), kitin terimobilisasi ditizon, atomic absorbtion spectrophotometer (AAS)
Article Details
License
Authors who publish with this journal agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).
PLAGIARISM CHECK
All manuscripts submitted to AJIE will be checked free of plagiarism elements. Manuscripts that do not meet the requirements will be returned to the author for further correction or will be rejected immediately.
The plagiarism check used by AJIE is Turnitin.